Accessories
Levelflex FMP56, FMP57 FOUNDATION Fieldbus
120
Hauser
16.1.3 Extension rod / centering HMP40
Accessory
Description
Extension rod / centering
HMP40
• can be used for:
FMP57
• Admissible
temperature
at lower nozzle edge:
– without center
washer:
no limitation
– with center washer:
-40 to 150 °C
(-40 to 302 °F)
• Additional information:
SD01002F
1
2
3
A0013597
1
Nozzle height
2
Extension rod
3
Center washer
010 Approval:
A
A: Non-hazardous area
M
M: FM DIP Cl.II Div.1 Gr.E-G N.I., zone 21,22
P
P: CSA DIP Cl.II Div.1 Gr.G + coal dust N.I.
S
S: FM Cl.I, II, III Div.1 Gr.A-G N.I., zone 0,1,2,20,21,22
U
U: CSA Cl.I, II, III Div.1 Gr.A-G N.I., zone 0,1,2
1
1: ATEX II 1G
2
2: ATEX II 1D
020 Extension rod; nozzle height:
1
115mm; 150-250mm / 6-10"
2
215mm; 250-350mm / 10-14"
3
315mm; 350-450mm / 14-18"
4
415mm; 450-550mm / 18-22"
9
Special version, TSP-no. to be spec.
030 Center washer:
A
Not selected
B
DN40 / 1-1/2", inside-d. = 40-45mm, PPS
C
DN50 / 2", inside-d. = 50-57mm, PPS
D
DN80 / 3", inside-d. = 80-85mm, PPS
E
DN80 / 3", inside-d. = 76-78mm, PPS
G
DN100 / 4", inside-d. = 100-110mm, PPS
H
DN150 / 6", inside-d. = 152-164mm, PPS
J
DN200 / 8", inside-d. = 210-215mm, PPS
K
DN250 / 10", inside-d. = 253-269mm, PPS
Y
Special version, TSP-no. to be spec.