Electrical connection
ISEmax compressed air cleaning unit (230 V AC)
12
Hauser
5.3
Post-installation check
‣
Once mounted, check all the connections to ensure they are secure and leak-tight.
‣
Check the cables and hoses for damage.
‣
Verify that the cables are routed in such a way that they are free from electromagnetic
interference influences.
6
Electrical connection
L
WARNING
Device is live!
Incorrect connection may result in injury or death!
‣
The electrical connection may be performed only by an electrical technician.
‣
The electrical technician must have read and understood these Operating Instructions and
must follow the instructions contained therein.
‣
Prior
to commencing connection work, ensure that no voltage is present on any cable.
6.1
Connecting the cleaning unit
NOTICE
Use of impermissible cable diameters!
IP65 degree of protection is voided
‣
Use cables with a diameter of 6 to 12 mm (0.24 to 0.47") for the Pg 13.5 cable gland.
‣
Use cables with a diameter of 13 to 18 mm (0.51 to 0.71") for the Pg 21 cable gland
(mains supply line).
NOTICE
Power consumption too high for the Liquiline alarm relay!
Can cause irreparable damage to the base module
‣
Connect the cleaning unit only to terminals of an additional module (AOR, 2R or 4R),
not
to the alarm relay of the base module.