Micropilot FMR60B HART
Operation options
Hauser
31
If an incorrect access code is entered, the user obtains the access rights of the
Operator
role.
7.4
Access to the operating menu via the local display
7.4.1
Device display (optional)
Possible to operate the optical operating keys through the cover. No need to open the
device.
Functions:
• Display of measured values and fault and notice messages
• Background lighting, which switches from green to red in the event of an error
• The device display can be removed for easier operation
Backlighting is switched on or off depending on the supply voltage and the current
consumption.
The device display is also optionally available with Bluetooth® wireless technology.
ESC
-
+
E
1
A0039284
8
Graphic display with optical operating keys (1)
• Key
• Navigate downwards in the picklist
• Edit the numerical values or characters within a function
• Key
• Navigate upwards in the picklist
• Edit the numerical values or characters within a function
• Key
• Change from main display to main menu
• Confirm entry
• Jump to the next item
• Select a menu item and activate the edit mode
• Unlock/lock the display operation
• Press and hold the key to display a short description of the selected parameter (if
available)
• key and key (ESC function)
• Exit edit mode for a parameter without saving the changed value
• Menu at a selection level: pressing the keys simultaneously takes the user back up a
level in the menu
• Press and hold the keys simultaneously to return to the upper level