Electrical connection
Source container FQG66
24
Hauser
6.2
Connecting the measuring device
6.2.1
Electrical connection
Required tools/accessories:
• Wire stripper
• If using stranded cables:
One ferrule for every wire to be connected
• Pliers for pressing on the ferrules
1 →
2 →
3 →
A0027173
A0027175
A0027176
‣
Release the 1/4 turn fasteners
on the cover of the terminal
housing.
‣
Fold down the cover of the
terminal housing.
‣
Guide the cable through the
slightly loosened cable entries.
To ensure tight sealing, do not
remove the sealing ring from
the cable entry.
4 →
5 →
6
8 (0.31)
8 (0.31)
A0027177
A0027178
A0027180
‣
Remove the cable sheath. Strip
the cable ends over a length of
8 mm (0.31 in). In the case of
stranded cables: also fit
ferrules.
‣
Firmly tighten the cable gland
and connect the cable
according to the terminal
assignment.
‣
Put the cover of the terminal
housing back on and fix it in
place with the 1/4 turn
fasteners.
6.2.2
Pneumatic connection
L
CAUTION
The throttle check valve is set at the factory and secured with thread locking fluid.
‣
Do not change the setting of the throttle check valve!
NOTICE
Pneumatic drive connection
‣
Compressed air connection: G1/8"
‣
The compressed air is connected to the check valve with choke.
6.3
Post-connection check
Are the device, cables and terminal housing undamaged (visual check)?