Prosonic M
7 Function group "extended calibr." (05)
Hauser
43
7.8
Function "offset" (057)
This function corrects the measured level by a constant value. The entered value is
added to the measured level.
7.9
Function "output damping" (058)
Influences the time an output requires to react to a sudden level jump (63% of steady
state). A high value attenuates, for example, the influences of rapid changes on the
measured variable.
User input:
0...255 s
The default value depends on the selected application parameters "
tank shape
"
(002)
,
"
medium property
"
(003)
and "
process cond.
"
(004)
.
7.10 Function "blocking dist." (059)
In this function the blocking distance is displayed. Level echoes within the blocking dis-
tance can not be detected by the Prosonic M. Make sure that the maximum level will
never run into the blocking distance.