![Endress+Hauser Ecograph T RSG30 Скачать руководство пользователя страница 59](http://html.mh-extra.com/html/endress-hauser/ecograph-t-rsg30/ecograph-t-rsg30_operating-instructions-manual_2408570059.webp)
Commissioning
59
Without affecting the internal memory, data packages are copied block-by-block to the CF card.
Tests are also made to determine whether the data has been written onto the data carrier without
any errors. The same happens when storing the data on the PC with the associated PC software.
!
Note!
• Before removing the CompactFlash card, select "CompactFlash card functions/Update CF". The
current data block is closed and saved to CompactFlash. This ensures that all current data (until
the last save) is included on the CF card.
• You are informed before the CF card is 100 % full. This is achieved by an acknowledgeable
message in the display, which indicates that the CF card described needs changing (only for "Stack
memory" ext. save mode, not for "Ring memory FIFO". ) In addition, a relay can be switched.
• Your unit knows which data has already been copied onto the CF card. Should you ever forget to
change it in time (or if there is no CF card inserted), the new CF card is filled with the missing
data from the internal memory - in as far as it is still available there.
• Since measured value acquisition/registration has highest priority, it can take some minutes until
the contents of the internal memory is copied onto the CF card.
• If the CF card is being accessed, the LED lights up. During this process, the CF card must not be
removed.
6.5.5
Main menu - Set up
Here you can trim your new unit to optimum operation, all possible operating parameters are
accessible (see see Section 6.4).
6.5.6
Main menu - Set contrast
You can set the viewing angle upwards or downwards to an optimum contrast depending on the
installation height.
6.5.7
Main menu - Diagnostic/unit information
Unit information and service functions for a swift unit check. The following functions are possible:
"CompactFlash (CF)
functions" menu items
Description
Safely remove CF
In order to ensure safe removal of the CompactFlash card all internal access is ended. You will receive a message when
the card can be safely removed.
!
Note!
Remove the card within 5 minutes otherwise the unit will start to automatically store data onto the card again. Only
remove the card using this function otherwise data can be lost!
Update CF
Data that has not been transferred to the CompactFlash card is now stored.
Complete memory to CF
The content of the internal memory is completely copied to the CompactFlash card. Please be patient! Measured value
recording continues in parallel and has the highest priority.
Copy set up to CF
All unit parameters (set up) are copied to the CompactFlash card. They can be archived and used for other units. The
file has the extension .rpd.
Load set up from CF
Loads unit parameters (set up) from the Compact Flash card into the internal nonvolatile memory of the unit. The file
has the extension .rpd.
Delete CF
Deletes all data on the CompactFlash card.
!
Note!
If a release code has been set up then the CompactFlash card can only be deleted after this code has been entered.