Installation
Deltabar S FMD77, FMD78, PMD75 with FOUNDATION Fieldbus
26
Hauser
4.3.7
Wall and pipe-mounting (optional)
Hauser offers the following mounting brackets for installing the device on pipes or
walls:
The standard mounting bracket version is not suitable for use in an application subject to
vibrations.
The vibration resistance of the reinforced version of the mounting bracket has been tested
according to IEC 61298-3, see the "Vibration resistance" section in the technical
documentation TI00382P.
When using a valve block, the block's dimensions must be taken into account.
Bracket for wall and pipe mounting including retaining bracket for pipe mounting and two
nuts.
material of the screws used to secure the device depend on the order code.
Technical data (e.g. dimensions or order numbers for screws) see accessory document
SD01553P/00/EN.
Please note the following when mounting:
• To prevent the mounting screws from scoring, lubricate them with a multi-purpose grease
prior to mounting.
• In the case of pipe mounting, the nuts on the bracket must be tightened uniformly with a
torque of at least 30 Nm (22.13 lbf ft).
• For installation purposes, only use the screws with item number (2) (see the following
diagram).
Standard design
Heavy duty design
A0031326
A0031327
Содержание Deltabar S FMD77