Deltabar PMD78B HART
Operation options
Hauser
37
7.4
Access to the operating menu via the local display
7.4.1
Device display (optional)
Functions:
• Display of measured values and fault and notice messages
• Background lighting, which switches from green to red in the event of an error
• The device display can be removed for easier operation
The device displays are available with the additional option of Bluetooth® wireless
technology.
Depending on the supply voltage and the current consumption, background lighting is
switched on or off.
Depending on the supply voltage and the current consumption, Bluetooth can
optionally be switched on or off.
1
4 5 6 7 8
2
3
A0047143
4
Segment display
1
Measured value
2
Bar graph proportional to the current output
3
Unit of measured value
4
Locked (symbol appears when the device is locked)
5
Bluetooth (symbol flashes if Bluetooth connection is active)
6
HART communication (symbol appears if HART communication is active)
7
Square root extraction (appears if a square root extraction of the measured value is output)
8
Measured value output in %
The following graphics are examples. The display depends on the display settings.