Description of device parameters
Deltabar FMD72
82
Example
• Current measured value: 0.3 ft
• New output unit: m
• New measured value: 0.3 m
Options
• %
• mm, cm, dm, m
• ft, in
• m
3
, in
3
• l, hl
• ft
3
• gal, Igal
• kg, t
• lb
Note
The unit selected is only used to describe the measured value. This means that when selecting a
new output unit, the measured value is not converted.
Factory setting
%
Height unit (026)
Navigation
Setup
®
Advanced Setup
®
Level
®
Height Unit (026)
Write permission
Operators/Service engineers/Expert
Description
Select the height unit. The measured pressure is converted to the selected height unit using the
"Adjust Density" parameter.
Prerequisite
"Level Selection" = "In height"
Options
• mm
• m
• in
• ft
Factory setting
m
Calibration mode (027)
Navigation
Setup
®
Advanced Setup
®
Level
®
Calibration Mode (027)
Write permission
Operators/Service engineers/Expert
Description
Select the calibration mode.