Deltabar FMD72
Description of device parameters
79
Factory setting
Cancel
Get URV (016)
Navigation
Setup
®
Advanced Setup
®
Current Output
®
Get URV (016)
Write permission
Operators/Service engineers/Expert
Description
Setting upper range value. The pressure for the upper current value (20 mA) is present at the device.
Use the "Confirm" option to assign the applied pressure value to the upper current value.
Prerequisite
Pressure measuring mode
Options
• Cancel
• Confirm
Factory setting
Cancel
Damping value (017)/(184)
Navigation
Setup
®
Damping value (017)/(184)
Write permission
Operators/Service engineers/Expert
(if the "Damping" DIP switch is set to "on")
Description
Enter damping time (time constant
t
) ("Damping" DIP switch set to "on")
Display damping time (time constant
t
) ("Damping" DIP switch set to "off").
The damping affects the speed at which the measured value reacts to changes in pressure.
Input range
0.0 to 999.0 s
Factory setting
2.0 sec. or according to order specifications
Level before lin. (019)
Navigation
Setup
®
Level Before Lin. (019)
Setup
®
Advanced Setup
®
Level
®
Level Before Lin. (019)
Diagnosis
®
Measured Values
®
Level Before Lin. (019)
Write permission
No write permissions. Parameter is read only.
Description
Displays the level value prior to linearization.