Deltabar FMD71, FMD72
Description of device parameters
63
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured pressure before the sensor trim.
Meas. press. HP (281)
Navigation
Expert → Measurement → Pressure → Meas. Press. HP (281)
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured HP pressure after sensor trim and simulation.
Meas. press. LP (282)
Navigation
Expert → Measurement → Pressure → Meas. Press. LP (282)
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured LP pressure after sensor trim and simulation.
Sensor Temp. LP (283)
Navigation
Expert → Measurement → Basic setup → Sensor Temp. LP (283)
Write permission
No write permissions. Parameter is read only.
Description
Displays the temperature currently measured in the sensor module. This can deviate from
the process temperature.
Sim. Press. HP (284)
Navigation
Expert → Diagnosis → Simulation → Sim. Press. HP (284)
Write permission
Operators/Service engineers/Expert
Description
Enter the simulation value. See also "Simulation Mode".
Prerequisite
"Simulation Mode" = Pressure HP
Value at switch-on
Current pressure measured value