Deltabar FMD71, FMD72
Description of device parameters
41
Description
Display of Device Revision (e.g. 1)
Sensor pressure HP (109)
Navigation
Expert → Measurement → Pressure → Sensor Press. HP (109)
Write permission
No write permissions. Parameter is read only.
Description
Displays the measured pressure before the sensor trim.
Sensor Temp. HP (110)
Navigation
Expert → Measurement → Basic setup → Sensor Temp. HP (110)
Write permission
No write permissions. Parameter is read only.
Description
Displays the temperature currently measured in the sensor module. This can deviate from
the process temperature.
Simulation mode (112)
Navigation
Expert → Diagnosis → Simulation → Simulation mode (112)
Write permission
Operators/Service engineers/Expert
Description
Switch on simulation and select the simulation mode. When changing the measuring
mode or the level type (Lin. Mode (037)) or when the device is restarted, any simulation
running is switched off.
Options
• None
• Differential pressure, → see this table, "Sim. pressure" parameter
• Level, → see this table, "Sim. level" parameter
• Press. HP, → see this table, "Sim. Press. HP" parameter
• Press. LP, → see this table, "Sim. Press. LP" parameter
• Tank content, → see this table, Sim. tank cont." parameter
• Current, → see this table, "Sim. Current" parameter
• Alarm/warning, → see this table, "Sim. error no."
Factory setting
None
Simulation
pressure LP
Measured
pressure LP
↓
↑
LP
→
Processing
Sensor LP measurement
→
+
Corrected
pressure
Simulation
differential
pressure