Deltabar FMD71, FMD72
Description of device parameters
57
Navigation
Expert → Measurement → Basic setup → Alarm cur.switch (165)
Write permission
No write permissions. Parameter is read only.
Description
Displays the switching state of DIP switch 3 "SW/Alarm min."
Display
• SW setting
The alarm current has the value defined in "Output Fail Mode" (051).
• Alarm min.
The alarm current is 3.6 mA, regardless of the software setting.
Corrected press. (172)
Navigation
Expert → Measurement → Pressure → Corrected press. (172)
Write permission
No write permissions. Parameter is read only.6
Description
Displays the measured differential pressure after position adjustment.
Tank description (173)
Navigation
Expert → Measurement → Linearization → Tank description (173)
Write permission
Operators/Service engineers/Expert
Description
Enter the tank description (max. 32 alphanumeric characters)
HART Input Stat. (179)
Navigation
Expert → Communication → HART Input → HART Input Stat. (179)
Write permission
No write permissions. Parameter is read only.
Description
Display of the HART input status
Bad / Uncertain / Good
Hart version (180)
Navigation
Expert → Communication → HART info → HART Version (180)
Write permission
No write permissions. Parameter is read only.
Содержание Deltabar FMD71
Страница 2: ......
Страница 68: ...www addresses endress com 71200289 71200289...