Electrical connection

Condumax CLS15D/16D/21D

20

Hauser

5.2  

Connecting the sensor

The electrical connection simulator to the transmitter is established using measuring cable

CYK10.

GN/YE

YE

GN

BN

WH

GND

+

Com A

Com B

GY

  A0024019

 10

Measuring cable CYK10

NOTICE

Mechanical twist protection for CLS15D and CLS21D

If too much force is applied to the Memosens head, this can shear the connections and

therefore destroy the sensor!

There is no need to exert excessive force when connecting the sensor to the cable coupling.

Proceed with care!

If the Memosens coupling clearly will not close, check the coupling for dirt or mechanical

damage, and make sure that you are turning it in the right direction. Pay attention to the

lock symbol on the coupling!

If necessary, use a different Memosens cable.

5.3  

Ensuring the degree of protection

Only the mechanical and electrical connections which are described in these instructions and

which are necessary for the required, designated use, may be carried out on the device

delivered.

Exercise care when carrying out the work.

Otherwise, the individual types of protection (Ingress Protection (IP), electrical safety, EMC

interference immunity) agreed for this product can no longer be guaranteed due, for example

to covers being left off or cable (ends) that are loose or insufficiently secured.

5.4  

Post-connection check

Device condition and specifications

Action

Are the sensor, assembly, or cables free from damage on the
outside?

Perform a visual inspection.

Electrical connection

Action

Are the mounted cables strain-relieved and not twisted?

Untwist the cables.

Is a sufficient length of the cable cores stripped, and are the
cores positioned in the terminal correctly?

Pull gently to check they are seated correctly.

Are all the screw terminals properly tightened?

Tighten the screw terminals.

Содержание CLS15D

Страница 1: ...roducts Solutions Services Operating Instructions Condumax CLS15D 16D 21D Sensors with Memosens protocol For conductive measurement of conductivity in liquids 0 BA01147C 07 EN 05 20 71464661 2020 01 3...

Страница 2: ...EU Declaration of Conformity Condumax CLS15D 16D 21D 2 Endress Hauser EU Declaration of Conformity ATEX NEPSI II 1G Ex ia IIC T3 T4 T6 Ga IECEx Ex ia IIC T3 T4 T6 Ga...

Страница 3: ...Condumax CLS15D 16D 21D EU Declaration of Conformity Endress Hauser 3 ATEX NEPSI II 3G Ex ic IIC T3 T4 T6 Gc...

Страница 4: ...ntification 10 3 3 Scope of delivery 11 3 4 Certificates and approvals 11 4 Installation 13 4 1 Mounting the sensor 13 4 2 Post installation check 17 5 Electrical connection 17 5 1 Quick wiring guide...

Страница 5: ...gerous situation Failure to avoid the dangerous situation can result in a fatal or serious injury LCAUTION Causes consequences If necessary Consequences of non compliance if applicable Corrective acti...

Страница 6: ...vity of liquids They are used in the following fields Sensor Applications Hazardous areas Condumax CLS15D Measurements in pure and ultrapure water Approved for Ex zone 0 Condumax CLS16D Measurements i...

Страница 7: ...certified conductivity sensors CLS15D CLS16D CLS21D may only be connected via the measuring cable CYK10 G I to the certified intrinsically safe digital sensor output circuits of measuring device Liqui...

Страница 8: ...to the certified intrinsically safe digital sensor output circuits of measuring device Liquiline M CM42 KV The electrical connection must be made according to the wiring diagram of the transmitter Me...

Страница 9: ...and product identification 3 1 Incoming acceptance 1 Verify that the packaging is undamaged Notify the supplier of any damage to the packaging Keep the damaged packaging until the issue has been resol...

Страница 10: ...For use in hazardous areas FM CSA IS NI Cl I Div 1 2 Gr A D Name Type Version Condumax CLS15D CLS16D CLS21D V V V Process connections materials not Ex relevant For use in hazardous areas ATEX NEPSI I...

Страница 11: ...on relating to your device is displayed in this window as well as the product documentation Manufacturer s address Endress Hauser Conducta GmbH Co KG Dieselstra e 24 D 70839 Gerlingen 3 3 Scope of del...

Страница 12: ...e Stating the individual cell constant 3 4 6 Biological reactivity test USP class VI CLS16D only Biological reactivity test certificate according to USP United States Pharmacopeia part 87 and part 88...

Страница 13: ...NPT or or clamp 1 As an option the sensor can also be installed using a commercially available T piece or cross fitting or using a flow assembly NOTICE Incorrect mounting or disassembly The Memosens h...

Страница 14: ...18 25 0 98 6 0 24 38 5 1 52 1 2 3 A0019014 2 With NPT thread in flow assembly 71042405 dimensions in mm inch 1 Sensor holder NPT 2 Inlet 3 Outlet 1 Ensure that the electrodes are fully immersed in th...

Страница 15: ...ork under air evacuated conditions Otherwise the CO2 in the air can dissolve in the water and its weak dissociation can increase the conductivity by up to 3 S cm 4 1 3 CLS21D Clamp connection Both she...

Страница 16: ...stalled via a flow assembly NOTICE Incorrect mounting or disassembly The Memosens head could become loose and fall off resulting in total sensor failure Only mount the sensor via the process connectio...

Страница 17: ...ost installation check 1 Are the sensor and cable undamaged 2 Is the sensor installed in the process connection and is not suspended from the cable 5 Electrical connection LWARNING Device is live Inco...

Страница 18: ...ne 0 IIC 2G CPS11D CPS12D CPS71D CPS72D CPS91D CPS92D CPS16D CPS76D CPS96D CPS171D IIC 1G IIC 3G EX Zone 1 Zone 0 Zone 2 CYK10 Ex ia Ga CPS41D CPS42D CYK10 Ex ia Ga EX CLS50D CLS15D CLS16D CLS21D COS2...

Страница 19: ...for zone 2 CPS11D CPS12D CPS71D CPS72D CPS91D CPS92D CPS16D CPS76D CPS96D CPS171D EX CYK10 Ex ic CPS41D CPS42D CYK10 Ex ic EX CLS50D CLS15D CLS16D CLS21D COS22D COS51D CPS471D CPS491D CPS441D CLS82D C...

Страница 20: ...he degree of protection Only the mechanical and electrical connections which are described in these instructions and which are necessary for the required designated use may be carried out on the devic...

Страница 21: ...efore applying pressure to an assembly with cleaning system ensure that the system has been connected correctly If you cannot reliably establish the correct connection do not install the assembly in t...

Страница 22: ...ldup from flue gas desulfurization or wastewater treatment plants Use a mixture of hydrochloric acid 3 and thiocarbamide commercially available and then rinse thoroughly with plenty of clear water 4 B...

Страница 23: ...ice for information on replacing the seal and recalibration in the factory 8 2 Return The product must be returned if repairs or a factory calibration are required or if the wrong product was ordered...

Страница 24: ...ion to water at 25 C 77 F CLS15D A 0 04 to 20 S cm CLS15D B L 0 10 to 200 S cm CLS16D 0 04 to 500 S cm CLS21D 10 S cm to 20 mS cm Temperature CLS15D 20 to 100 C 4 to 212 F CLS16D 5 to 100 C 23 to 212...

Страница 25: ...asuring system traceable to NIST or PTB The exact cell constant is entered into the manufacturer inspection certificate supplied The uncertainty of measurement in determining the cell constant is 1 0...

Страница 26: ...operation 5 to 120 C 23 to 248 F Sterilization max 45 min Max 150 C 302 F at 6 bar 87 psi absolute CLS21D 20 to 135 C 4 to 275 F at 3 5 bar 50 psi absolute 1 Threaded versions max 30 minutes The maxi...

Страница 27: ...20 40 60 80 100 120 140 C 212 p abs T A0031428 EN 11 Mechanical pressure temperature resistance A Can be sterilized for a short time 1 hour CLS16D 6 5 20 60 100 120 150 23 60 68 120 180 240 248 300 20...

Страница 28: ...anical construction 9 5 1 Weight CLS15D and CLS21D Approx 0 3 kg 0 66 lbs depending on version CLS16D Approx 0 13 to 0 75 kg 0 29 to 1 65 lbs depending on version 9 5 2 Materials CLS15D Electrodes Pol...

Страница 29: ...eel 1 4435 EPDM 9 5 3 Process connection CLS15D Thread NPT and Clamp 1 as per ISO 2852 CLS16D Clamp 1 1 2 as per ISO 2852 also suitable for TRI CLAMP DIN 32676 Tuchenhagen VARIVENT N DN 50 to 125 NEUM...

Страница 30: ......

Страница 31: ......

Страница 32: ...www addresses endress com 71464661 71464661...

Отзывы: