Installation
Cleanfit P CPA472D
8
Hauser
3
Installation
3.1
Incoming acceptance, transport, storage
• Make sure the packaging is undamaged!
Inform the supplier about damage to the packaging. Keep the damaged packaging until the matter
has been settled.
• Make sure the contents are undamaged!
Inform the supplier about damage to the delivery contents. Keep the damaged products until the
matter has been settled.
• Check that the scope of delivery is complete and agrees with your order and the shipping
documents.
• The packaging material used to store or to transport the product must provide shock protection
and humidity protection. The original packaging offers the best protection. Also, keep to the
approved ambient conditions (see "Technical data").
• If you have any questions, please contact your supplier or your sales center responsible.
3.2
Installation conditions
3.2.1
Notes on installation
The assembly is designed for installation on tanks and pipes. Suitable nozzles must be available for
this.
When using standard glass electrodes, only installation positions are permitted in which the middle
axis of the assembly lies at least at an angle of 15° from the horizontal (see diagram). Otherwise,
there will not be a reliable contact between the inner side of the pH membrane and the inner
terminal leads via the electrolytes.
When using an ISFET Tophit sensor, there are, in principle, no restrictions for the installation
position. An installation angle of 0 to 180° is, however, recommended.
a0005327
Fig. 2:
Installation angle
A
Glass electrodes: minimum 15 ° to the horizontal
B
ISFET sensors: recommended 0 ... 180 °, overhead possible
15°
15°
A
B