9.3
Process faults without messages
9.4
Spare parts
The measurement station is constructed using a number of individual instruments. An
itemised parts list is made for each measurement station and placed in the device ring
binder. This means that the user can quickly identify individual items.
The parts list can be found in section 3 in the ring binder.
However, it is recommended that the user holds the most important consumables and
parts subject to wear and tear on stock (see maintenance plan in section 10 in the ring
binder). In order to make the correct decission the storage time and conditions must be
considered, e.g. pH sensors etc.
Fault
Possible cause
Solution
No function
•
No power supply
•
Main isolator, fused switches
tripped
•
Call skilled electrician
Pump does not run or runs only
intermittently
•
Dry run protection active
•
Thermal overload or short
circuit on the motor
•
No medium available
•
Call skilled electrician
Feed rate of the eccentric pump
reduces
•
Suction hose is partially
blocked
•
Stator and rotor are worn
•
Check suction part and
pipes and if required clean or
change
•
See op. manual for the
eccentric pump (section 8 in
the ring binder)
Cleaning system does not
operate
•
No water available
•
No voltage on the controller
•
Cleaning system not
switched on
•
Pipework over pressure
monitor active. Blockage in
the pipwork after the pump
•
No medium available
•
Pump does not run
•
Check water connection
•
Check electrical
connections, see electrical
connection schematic in
section 3 in the ring binder
•
Set selector switch "Rinse
ON/OFF" in swing frame to
"ON"
•
Remove blockage from
pipework
•
Dry run protection is active.
Check suction hose and
hose end for blockages
•
Check pump motor, check
pump for blockages; pump is
not switched on