12.2 Performance characteristics
12.2.1 Reference operating conditions
Temperature
20 °C (68 °F)
pH value
pH 6 to 7
Flow
40 to 60 cm/s
ClO
2
-free base medium
Deionized water
12.2.2 Response time
T
90
< 15 s (after completing polarization)
12.2.3 Measured value resolution of sensor
CCS50D-**11AD
0.03 µg/l (ppb) ClO
2
CCS50D-**11BF
0.13 µg/l (ppb) ClO
2
CCS50D-**11CJ
1.1 µg/l (ppb) ClO
2
12.2.4 Measured error
±2 % or ±5 µg/l (ppb) of value measured (depending on which value is higher)
LOD (limit of detection)
LOQ (limit of quantification)
CCS50D-**11AD
0.0007 mg/l (ppm)
0.002 mg/l (ppm)
CCS50D-**11BF
0.0013 mg/l (ppm)
0.004 mg/l (ppm)
CCS50D-**11CJ
0.0083 mg/l (ppm)
0.025 mg/l (ppm)
12.2.5 Repeatability
CCS50D-**11AD
0.002 mg/l (ppm)
CCS50D-**11BF
0.007 mg/l (ppm)
CCS50D-**11CJ
0.025 mg/l (ppm)
12.2.6 Nominal slope
CCS50D-**11AD
195 nA per 1 mg/l (ppm) ClO
2
CCS50D-**11BF
50 nA per 1 mg/l (ppm) ClO
2
CCS50D-**11CJ
6 nA per 1 mg/l (ppm) ClO
2
4)
Based on ISO 15839. The measured error includes all the uncertainties of the sensor and transmitter (measuring
chain). It does not contain all the uncertainties caused by the reference material and adjustments that may have
been performed.