Product description
CCS50D
8
Hauser
3
Product description
3.1
Product design
The sensor consists of the following functional units:
• Membrane cap (measuring chamber with membrane)
– Separates the inner amperometric system from the medium
– With robust PVDF membrane and pressure relief valve
– Guarantees a defined and consistent electrolyte film between cathode and membrane
– With special support grid between cathode and membrane for a defined and consistent
electrolyte film and therefore a relatively constant reading at varying pressures and flows
• Sensor shaft with
– Large anode
– Cathode embedded in plastic
– Embedded temperature sensor
3
1
2
4
5
7
6
8
A0034227
1
Sensor structure
1
Memosens plug-in head
2
Sensor shaft
3
Large anode, silver/silver chloride
4
Gold cathode
5
Grooves for installation adapter
6
Membrane cap with dirt-repellent
membrane
7
Pressure relief valve (elastic)
8
Sensor membrane
3.1.1
Measuring principle
Chlorine dioxide levels are determined in accordance with the amperometric measuring
principle.
The chlorine dioxide (ClO
2
) in the medium diffuses through the sensor membrane and is
reduced to chloride ions (Cl
-
) at the gold cathode. At the silver anode, silver is oxidized to