Commissioning
Liquiline System CA80COD
44
Hauser
Hose connection diagram
R
P
D
DU
DM
W
E
RK
S1
S0
RB
RN
A0028827
41
Liquiline System CA80COD
P
Sample
D
Procedure
DM
Dilution module (optional)
S0
Zero standard 0
W
Water for optional dilution module
RN
Reagent RN
RK
Reagent RK
RB
Reagent RB
S1
Standard 1
E
Ventilation
DU
Dosing system
R
Pressure reactor
7.2
Function check
L
WARNING
Incorrect connection, incorrect supply voltage
Safety risks for staff and device malfunctions
‣
Check that all connections have been established correctly in accordance with the wiring
diagram.
‣
Ensure that the supply voltage matches the voltage indicated on the nameplate.
‣
Ensure that the reactor's safety cover is undamaged and installed correctly.
Before putting the device into operation:
‣
Connect the liquid-bearing hoses of the sample supply system.
• After mounting, check all the connections to ensure they are secure and leak-tight.
• Visually inspect all the hose connections to ensure everything is correct.