Wiring
ASP station 2000 peristaltic
78
Hauser
4.1.3
Terminal position
Fig. 5:
Electronics compartment after mounting the cover (
A: Position of the shielded terminals in electronics compartment
B: Position of the terminal block in the electronics compartment
C: Position of the battery for sampling charge buffer operation
4.1.4
Terminal block connection
Fig. 6:
Assignment and wiring diagram of the terminal block
!
Note!
Use a screened cable for the signals. The screen is connected to the shielded terminals provided for
this purpose. If possible, the screen should be laid on both sides.
You can set the following signals at the signal cable:
• Input signals:
- 2 digital signals > 20 ms
- 1 analogue signal 0/4...20 mA
• Output signals:
- 2 digital signals
A detailed description of the signal assignment (inputs and outputs) is provided in the Section
"Device configuration".
L1
N
PE
L
N
Power
PE
230 VAC / 50 Hz
12
11
10
9
8
7
6
5
4
3
2
1
+
-
-
+
R
l
-
+
R
l
-
+
R
l
-
+
-
+
+
-
-
+
R
R
l
l
+
-
white
brown
green
yellow
grey
pink
blue
red
black
violet
grey/pink
red/blue
0 to 20 mA
4 to 20 mA
10 V max. 30 mA load!
Gnd
U+
Input 1 +
Input 1 Gnd
Input 2 +
Input 2 Gnd
Output 1 +
Output 1 -
Analogue -
An
int. Shunt: 50 Ohm
>20 ms
Only low voltage!
<28 V
max. current when using
ext. voltage
Imax=100 mA
Output 2 +
Output 2 -
Analogue input
Power supply
Digital in (DI)
Outputs
Shield
Содержание ASP Station 2000 peristaltic
Страница 66: ...ASP Station 2000 Peristaltik Index 66 Endress Hauser W Warenannahme 8 Werkstoffe 63 ...
Страница 197: ...ASP Station 2000 péristaltique Index Endress Hauser 197 ...
Страница 198: ...Endress Hauser ...
Страница 199: ...Endress Hauser ...
Страница 200: ...www endress com worldwide BA176R 09 a3 04 06 51008239 FM SGML6 0 ProMoDo ...