ASP station 2000 peristaltic
Installation
Hauser
75
3.2.3
Storage
Observe the following points:
• The device must be packed in shockproof packaging for storage (and transport). The original
packaging offers the best protection for this.
• The permitted storage temperature is -20...+60 C (pref20 C).
3.3
Mounting conditions
3.3.1
Dimensions
You can find the dimensions of the sampler in Section 10 "Technical data - Mechanical
construction."
3.3.2
Mounting location
Fig. 3:
Notes on mounting the device
When mounting the device, please observe the following points:
• Position the device on a level surface.
• Protect the device against additional heat (e.g. radiators).
• Protect the device against mechanical vibration.
• Protect the device against strong magnetic fields.
• Ensure that air can circulate unhindered at the back of the cabinet. Do not position the device
directly against a wall (the distance between the wall and the rear side of the cabinet should be
at least 100 mm).
• Do not position the cabinet directly above the channel of the inflow to the wastewater treatment
plant (sulphurous vapours!)