ASP Inline
Wiring
Hauser
39
4.1.2
Other control units
Fig. 8:
Electrical connection of the ASP Inline sampler assembly with options "rinse water and compressed air" to a
PLC control unit
4.1.3
Power supply connection
"
Caution!
• Before wiring the device, ensure that supply voltage corresponds to the specification on the
nameplate.
4.2
Degree of protection
Valve/terminal housing: IP65
4.3
Post-connection check
After the electrical installation of the assembly, perform the following checks:
S1
U
-
+
L1
L2
43
41
42
41
42
43
S2
S3
X1
X2
S4
S5
2
1
L1: Signal “cylinder down"
L2: Signal “cylinder up”
U : 24 VDC power supply
S1: Valve control
S3: Reed contact “up”
S2: Reed contact "down”
bk
rd
bk
rd
S5: Valve - rinse water
S4: Valve - compressed air
Valve
sampling
Option
valve
compr. air
Option
valve
rinse water
Device status and specifications
Notes
Is the assembly or the cable damaged (visual inspection)?
-
Electrical connection
Notes
Does the supply voltage match the information on the nameplate?
24 V DC
Is the wiring correct?
-
Are the mounted cables relieved of tension?
-
Are the power supply and signal cables connected correctly?
see wiring diagram on the
housing
Are all of the terminals tighten?
-