Installation
ASP Inline
34
Hauser
3
Installation
3.1
Installation conditions
The ASP Inline sampler assembly is suitable for installation on pressurised pipelines with a nominal
diameter of minimum DN50. Installation is performed using a flange connection.
The permitted ambient temperature (see "Technical Data" chapter) must be maintained during
installation and operation.
3.1.1
Dimensions
The dimensions of the ASP Inline sampler assembly (5 ml and 50 ml sample volume) can be found
in the "Technical Data" chapter.
3.1.2
Installation site
Fig. 2:
Installation position of the ASP Inline sampler assembly
Item A: Do not install at the highest point of the pipe - risk of air build-up!
Item B: Do not install in a down pipe leg.
Item C: The sampler assembly must be positioned in a permitted mounting location above the sampling device or the
sampling vessel.
Item D: Do not install the ASP Inline sampler assembly below the sampling device or the sampling vessel (e.g. ASP Station
2000).
B
A
C
ASP-Station 2000
ASP-Station 2000
D