ZEEnit 700 Q
Installation and initial start-up
15
3
Installation and initial start-up
The device may only be assembled, installed and repaired by service engineers from
Analytik Jena or by technical personnel authorized by Analytik Jena.
When installing and starting up your machine, please observe the safety instructions in
Section "Safety instructions". Compliance with these safety instructions is a require-
ment for error free installation and the proper functioning of your AAS measuring
environment. Always observe all warnings and attention messages which are dis-
played on the device itself or which are displayed by the control program of the ZEEnit
700 Q. The customer must ensure that the installation conditions at the place of in-
stallation comply with the specifications stipulated by Analytik Jena.
3.1
Starting up the ZEEnit 700 Q with accessories
3.1.1
Switching on sequence, daily work commencement
1.
Switch on the PC and wait for the computer program to initialize: The user icons
appear on the screen, including the ASpect LS program icon.
2.
Switch on the ZEEnit 700 Q: Press the green ON/OFF switch on the right side wall.
3.
Start the ASpect LS program: Double-click on the ASpect LS symbol.
4.
Connect the printer and the compressor if they are needed.
The AAS system is now switched on, work may begin.
NOTICE
The mobile cooling unit KM5 is controlled by the ZEEnit 700 and is therefore not
switched on manually.
3.1.2
Switching off sequence
1.
On the PC close the application program Aspect LS: Click the F
ILE
▸
C
LOSE
menu
options.
2.
For unsaved values decide whether unsaved data/information should be saved
before exiting the program.
3.
Shut down the PC.
4.
Use the respective mains switches to switch off (in this order):
Compressor
AAS accessories
AAS
Printer and PC
The AAS system is now switched off.