CyBio Carry
Technical description
17
Name
Function
On / Off
On/Off switch
24 V DC, 4 A
Connection for external power supply unit
(with specification of the operating voltage
and maximum current consumption)
ESTOP 1
E-STOP OUT, forwarding interface to an-
other device
ESTOP 2
E-STOP IN, connection for an opening switch
element (STOP function)
STOP
Connection (2 x 24 V DC, 0.1 mA) for an
opening switch element (STOP function)
BC
RS 232 interface for connecting a barcode
reader
ETH
Ethernet port for network connection
COM 1
RS 232 interface for connection to the con-
trol computer
COM 2
RS 232 forwarding interface for connecting
additional lab devices using the same control
computer
External power supply unit
The power supply unit is a wide range power supply unit. The voltage supply on the pri-
mary side comes from a power socket providing an AC voltage of 100 to 240 V ± 10 %
and a frequency of 50/60 Hz.
On the secondary side, the power supply unit provides a DC voltage of 24 V (max.
3.75 A).
1
2
Fig. 11
External power supply unit
1 Input
2 Output
3.3
Labware requirements
The labware is usually microplates. The labware must comply with the ANSI/
SLAS 1-2004 (Microplates - Footprint Dimensions) and ANSI/SLAS 2-2004 (Mi-
croplates - Height Dimensions) standards:
¡
The labware must be rectangular.
¡
The labware must meet the minimum width and minimum depth requirements.
¡
The labware must have a circumferential edge.