Operation
Biometra TOne
34
Graphical programming
The currently selected step is highlighted in blue during the programming process. The
graphical view divides each step into heating phase and holding time. Some parameters
are only displayed by symbols (see table).
Name:
Lid:
Preheat Lid:
On
Home
Edit
Insert
Delete
Done
Back
Step
Go To
Cycles
∆T
∆t
∆R
30x
1
2
3
4
5
6
Step
Step
Step
Step
Step
Go To
Temperature [°C]
Three-step
dd.mm.yyyy hh:mm
Program Runtime:
1h 15m 10s
User:
John Doe
Fig. 13
Graphical programming
Program parameters
Display
(example)
Description
Name:
Three-step
Define the program name
Lid:
99 °C
Define the temperature of the heated lid
(30 to 110 °C)
Preheat Lid:
On
Enable or disable the preheating of the
heated lid before the actual PCR run
Step
Step
1
Define at least one block temperature and
holding time for each program step. All other
parameters are optional.
For more than 6 program steps, use the ar-
row keys to navigate through the displayed
program
Temperature:
Hold Time (h:mm:ss)
95.0
0:05:00
Define a block temperature
(3 to 99 °C ±0.1 °C) for each step
Define a holding time during which the block
temperature is maintained
(≤9 h 59 min 59 s)
Use
∞
to program a pause
GoTo
Cycles
Cycles 30x
Define program loops:
¡
Enter the step number to which the pro-
gram shall return
¡
Define the number of iteration cycles
(max. 999)
ΔT
ΔT
(in red)
Increase or decrease the block temperature
in each cycle by the temperature increment
(
Temp. incr./decr.:
) of ±0.1 to 20 °C
Δt
Δt
(in red)
Extend the holding time in each cycle by the
time increment of 1 to 240 s