Maintenance
TOPwave
66
Start the application and follow the program instructions.
When the program is completed the new calculated correction value will be displayed.
NOTICE
The adjustment can be repeated any number of times with any number of vessels. How-
ever, at least 2 vessels must be in the oven. If necessary a vessel with a standard coup-
ling cap must be used as the second vessel.
8.2
Output power measurement
The output power of the magnetron has been thoroughly tested and certified before it is
delivered. However all magnetrons are subject to an aging process and it may become
necessary to check the output power. This is definitely advisable if the rate at which it
heats becomes significantly slower.
To measure the output power 500 ml water is heated at full microwave power for 42
seconds and then the rise in temperature established. The output power can be calcula-
ted from this
You need a thermometer for the measuring the water temperature bevor and after the
performance test.
Fill in a Teflon/PP/glass beaker 500 ml water.
Load the application of the power test.
Press the button [A
PPLICATION
] in the Main menu.
Press the button [power test] in the application library.
The main menu appears.
Press the button [
START
].
Follow the program instructions and quit each step with the function buttons.
The output power should be at least 900 watts.
If values of less than 800 watts are determined, the Analytic Jena Support should be
contacted.
Principle of power measu-
rement
Implementation of power
measurement