PlasmaQuant MS Series
Maintenance
Forms and Checklists
213
15.2
Operator Checklist
Routine Maintenance – Operator Checklist
Copy this page, complete the checklist and insert it into the instrument logbook.
Check
Component
Action
Frequency
Torch
Check for verification and surface coating
of sample or carbon. Clean if required.
Weekly or as required depending
on sample matrix
Transfer tubing
Check for coating or deposits on the
internal surfaces, and clean if required.
Every three months or as required
depending on sample matrix
Sample and skimmer
cones
Clean using stainless steel cleaning
powder then in an ultrasonic bath with a
caustic detergent followed by de-ionized
water.
Weekly or as required depending
on sample matrix
Extraction lens #1 and #2
Clean using stainless steel cleaning
powder then in an ultrasonic bath,
followed by de-ionized water.
Every 3 months or as required
depending on sample matrix
Rotary pump
Check the oil level and color and change
oil as needed.
Weekly
Water cooler
Inspect water level in cooler. Add
additional water if necessary.
Weekly
Spray chamber
Check for coating or deposits on the
internal surfaces, and clean if required.
Weekly or as required depending
on sample matrix
Peristaltic pump tubing
Replace the tubing as required.
Weekly or as required depending
operational hours and sample
matrix
Nebulizer
Clean using the instructions outlined in
the Aspect MS Help.
Weekly or as required depending
on sample matrix
Operators Name: _______________________ Signature: ___________________________