multi N/C 2100S (duo, pharma)
Operation
55
}
Open the Measurement |Calibration menu option.
}
In the query that follows, select either the method to calibrate or a preexisting cali-
bration table.
}
Carry out the software instructions that follow.
ü
The Calibration - Data of new calibration window opens.
Figure 26Data for new calibration
}
Select the calibration type in the Calibration Parameters group:
– Preferably, perform Multiple point calibrations with a
constant sample volume
and variable concentrations. The volume set in the method is entered automati-
cally in the constant sample volume input field. Only change the presets if the in-
jection volume of the standard solutions differs from the volume set in the
method.
}
For calibrations with a
constant concentration
, enter the concentration of the stan-
dard solution in the input field.
}
Enter the number of calibration points in the Number of standards input field.
}
Select the analysis parameters of the loaded method to calibrate in the Analysis pa-
rameters field.
– Activate the IC and TC parameters individually for calibration of the NPOC plus
parameter and higher concentrations of c > 0.5 mg/l.
}
Use the calibration for the TOC/NPOC parameter for the lower concentration range.
A single point calibration is sufficient in the lower concentration range.
}
The software displays the sample feed type under Sample introduction. The indica-
tion is for information only and cannot be modified.
}
Select how to take the blank value of the prepared water of the standard solutions
into account in the Preparation blank group
– Measure selection field:
The content of the preparation water is measured separately immediately before
calibration. To do this, place a vessel with preparation water at the first position
on the autosampler.
With manual sample supply, the software sends a prompt to provide the prepara-
tion water.