multi N/C 2100S (duo, pharma)
Installation and commissioning
47
4.3.3
External solids module
NOTICE
Observe accessory instructions
This accessory has separate instructions containing important information and measures
for hazard prevention.
¡
Observe the separate instructions for the accessory during installation.
Installation of the modular multi N/C 2100S duo measurement system for automated
solids analysis is described in the separate operating manual for the HT 1300 solids
module.
Connection to the analyzer
}
Set up the solids module next to the analyzer.
}
Connect the "analyte" connection on the solids module to the "analyte" connection on
the rear of the analyzer.
}
Connect the "pump" connection on the solids module to the "CLD/pump" connection
on the rear of the analyzer.
}
Connect the connection hose for oxygen to the gas supply pressure reducer and to
the "oxygen" gas connection on the rear of the solids module. Set an inlet pressure of
200 to 400 kPa on the pressure reducer.
}
Connect the supplied serial data cable to the "CLD/HT" connection on the rear of the
analyzer. Connect the other end of the data cable to the solids module.
}
Start the multiWin program. Switch on the components of the analysis system. Ini-
tialize the analysis system.
}
Open the Options window with menu option Configuration | Edit options. Go to the
Analyzer components tab.
}
Activate the External solids module option.
}
Exit the window with [OK].
1
2
3
4
5
6
Figure 24Connections on the backplate of the solids module
1 Analyzer interface
2 Mains connection
3 Measuring gas outlet "OUT"
4 Oxygen inlet "O
2
"
5 Pump connection "pump"
6 Measuring gas connection "analyte"