Installation and commissioning
multi EA 5100
30
5
Installation and commissioning
5.1
Installation conditions
5.2
Installation location requirements
Ambient conditions
172). If required, ensure that the room is
temperature-controlled.
Laboratory conditions
The device is only approved for indoor use. The installation location should have the
characteristics of a chemical laboratory. It must meet the following conditions:
¡
Atmosphere free from hydrocarbons, halogens, sulfur compounds and nitrogen ox-
ides
¡
Atmosphere with low dust levels
¡
No vibrations
¡
No smoking in the operating room of the device
Installation location require-
ments
The requirements for the installation location of the device are as follows:
¡
No caustic vapor in the immediate vicinity of the device and its system components.
These could corrode the device connections and modules.
¡
Free from draft air; do not install the device close to windows or doors
¡
Away from electromagnetic sources of interference
¡
No direct sunlight and away from radiant heaters
¡
The front door and air vents must not be obstructed by other equipment or furnish-
ings
¡
Maintain a safety distance of at least 20 cm from the rear of the device to other de-
vices or walls.
5.3
Power supply
WARNING
Danger due to electrical voltage
¡
Only connect the device to a properly grounded socket which complies with the volt-
age indicated on the device’s rating plate.
¡
Do not use an adapter in the feeder.
The device operates on single-phase alternating current.
The installation of the electrical equipment in the laboratory must comply with the
DIN VDE 0100 standard. At the connection point, an electrical current in accordance
with the standard IEC 38 must be available.
The electrical connection data can be found in the technical data (→ "Technical data for
the multi EA 5100 "
172).