Maintenance and care
multi EA 5000
150
}
Switch off the detection module via the power switch.
}
Remove the left side panel. Remove the 4 screws to do this. Remove the protective
grounding cable and remove the side panel.
}
Disconnect the following hose connections:
Disconnect hose 25 from the T piece.
Unscrew hose 24 from the bottom of the ozone decomposers.
}
Remove the ozone decomposer with its filter and hose 25 from the holding clamps.
Recommendation: Remove it from the top clamp first.
}
Install the new filter in the reverse order.
ü
The detection module is ready for measurement.
1
2
3
Figure 70Replacing the chemical ozone decomposer
1 Ozone decomposer
2 Hose 24
3 Filter with hose 25