Troubleshooting
multi EA 5000
116
Error
Possible cause
Remedy
Smell of ozone
¡
Chemical ozone de-
composer old/inef-
fective
¡
Replace chemical
ozone decomposer
(→ "Replacing the
chemical ozone de-
composer"
¡
Gas hoses on the
ozone generator
leaky or loose
¡
Check hoses
¡
Search for leak as
necessary with indi-
cator paper
¡
Device leaking or de-
fective
¡
Inform customer ser-
vice department
13.8 Analytical problems during TC determination
Error
Possible cause
Remedy
Scattering measurements
¡
Oxygen supply to the
module interrupted
¡
Check oxygen con-
nection, reconnect as
necessary
¡
Absorber clogged or
worn out
¡
Replace absorber,
see (→ "Replacing the
absorber"
¡
Sample unsuitable for
vertical operation
mode (droplet forma-
tion)
¡
Use the horizontal
operation mode
¡
In vertical operation
mode: Quartz wool
not present or at the
wrong position in the
combustion tube
¡
Check the position of
the quartz wool, ad-
just as necessary
¡
Sample evaporates
before dosing
¡
Use a cooled sampler.
¡
Sample too viscous
too be drawn in with-
out air pockets
¡
Use the horizontal
operation mode and
dilute the sample, or
dose directly as a
solid.
¡
Inhomogeneous sam-
ple or sample con-
taining particles
¡
Homogenize sample
Very low measurement
values or no analysis sig-
nal
¡
Measuring gas con-
nection incorrect
¡
Check the measuring
gas connection
¡
Gas outlet flow im-
peded
¡
Check the "sample
out" connection for
unimpeded flow
¡
Oxygen supply inter-
rupted
¡
Check oxygen con-
nection, reconnect as
necessary