Maintenance and care
multi EA 4000
114
7.3
Operation with Cl module
Maintenance task
Maintenance interval
Remove splashes of chemicals from the
housing
Immediately
Inspect the combustion tube and replace
if necessary
After approx. 100 operating hours
Clean measuring cell
At least weekly
Recommendation: wipe clean after electrolyte
replacement
Replace electrolyte solution
As required, min. once per measuring day
Sulfuric acid replacement
If the sulfuric acid drops below 70 %, i.e. the
acid volume increases by 1/3rd.
Replace the halide trap
when half of the copper wool or the brass wool
is discolored
Replacing the adsorption agent
Weekly (every 40 operating hours)
The copper and brass wool in the halide trap is replaced analogous to the replacement
of the halide trap in the basic device multi EA 4000 (
→
see section "Replace the halide
7.3.1
Replacing the sulfuric acid/cleaning the sulfuric acid container
WARNING
The concentrated sulfuric acid used as desiccant in the sulfuric acid container can cause
serious burns!
Wear appropriate protective clothing when working at the sulfuric acid container.
1.
Pull the gas olive (1) out of the sulfuric acid container.
2.
When using a container with split outlet:
Pull the black hose (3) off the bypass of the sulfuric acid
container.
3.
Hold the sulfuric acid container, remove the forced clamp
(2) and detach the spherical joint connection.
4.
Carefully take the sulfuric acid container out of the drip
tray.
5.
Dispose of the sulfuric acid. Observe the regulations for
handling concentrated sulfuric acid.
Maintenance intervals
Содержание Analytik Jena multi EA 4000
Страница 1: ...Operating Manual multi EA 4000 Elementary analyzer C S and Cl solids analysis...
Страница 50: ...Commissioning multi EA 4000 50 Figure 32 multi EA 4000 C S Hose diagram...
Страница 52: ...Commissioning multi EA 4000 52 Figure 33 multi EA 4000 Cl Hose diagram...
Страница 54: ...Commissioning multi EA 4000 54 Figure 34 multi EA C S Cl for C S measurements Hose diagrams...
Страница 55: ...multi EA 4000 Commissioning 55 Figure 35 multi EA 4000 C S Cl for Cl measurements...
Страница 57: ...multi EA 4000 Commissioning 57 Figure 36 Automatic TIC solids module Hose diagrams...
Страница 58: ...Commissioning multi EA 4000 58 Figure 37 multi EA 4000 C S with automatic TIC solids module...
Страница 60: ...Commissioning multi EA 4000 60 Figure 38 multi EA 4000 with man TIC solids module Hose diagrams...
Страница 61: ...multi EA 4000 Commissioning 61 Figure 39 multi EA 4000 C S Cl with man TIC solids module...