InnuPure C16
touch
Safety instructions
9
The instruction also includes imparting the contents of this user manual. Furthermore,
the user manual must be accessible to the operating and service personnel at any
time!
2.5
Safety instructions, transport and installation
Only transport the device in its original packaging! Ensure that all transport protec-
tions have been fitted and the device is completely empty, cleaned and decontami-
nated.
The InnuPure C16
touch
may only be installed by the service personnel of Analytik
Jena or by persons authorized by Analytik Jena.
To prevent health damage the following must be observed when moving the device in
the laboratory (lifting and carrying):
For reasons of safety 2 persons are required to transport the device and must posi-
tion themselves on both sides of the equipment.
Since the device does not have handles, grip the device firmly with both hands at
the lower end, lifting it simultaneously. Make sure the door is closed. Do not hold
onto the door area or the plexiglas front.
The guide values and statutory limits for lifting and carrying loads without auxiliary
equipment must be observed and adhered to.
2.6
Safety instructions - operation
2.6.1
General
The device may only be operated if all protective equipment (front doors) are in
place, properly installed and fully operational.
Free access to the device switch on the back of the enclosure must be ensured dur-
ing operation.
During operation and after aborting an extraction protocol, the operator must not
reach with his hands into the device chamber. There is a risk of crushing due to the
motor-driven carriage.
Careless handling during preparation of the sample tray may lead to crushing of
fingers when the clamps for locking of the reagent plates and strips are pushed
down.
The ventilation equipment on the device must be in good working condition. Cov-
ered ventilation grilles or slits etc. may cause the device to break down or may
cause damage to it.
Do not introduce any objects into the device and prevent liquids from entering the
device through openings or gaps.
The operating system of the tablet and the IP
extract
program form a closed system.
There is no active antivirus protection. The user is responsible for data safety of in-
serted USB Memory Sticks.