Specification
11.1
Technical data
System parameters
Extraction principle
Based on surface-functionalized magnetic or paramagnetic parti-
cles or based on the Smart extraction principle
Extraction duration
< 45 min (without lysis)
< 100 min (with lysis)
MAG kits yield
Depending on the sample type and quantity
Tissue (20 mg): up to 50 µg
Whole blood (200 µl): up to 10 µg
Plant (100 mg): up to 60 µg
SE kits yield
Depending on the sample type and quantity
Tissue (5-20 mg): between 15 and 150 µg
Whole blood (0.5-3 ml): between 5 and 90 µg
Drive
Low-noise long-life stepper and servomotors
(noise emission max. 65 dB)
Device operation
Stand-alone via IP
extract
with 10”
touch
screen (color)
Maintenance
Maintenance-free through the use of non-wearing stainless steel
pistons
Cleaning
System components are easily accessible through large, split
front door
Temperature control
Heating up to 90 °C
Application parameters
Consumables
Completely contained in the required kit
Sealed, pre-filled reagent strips or plates
Lysis step
Automated in the device (dependent on the source material)
Extraction routines
Pre-installed protocols (optimized for the most varied source ma-
terials)
Cross contamination
Completely excluded through sealed reagent strips and plates
Tips
Max. 32 per preparation each with a capacity of 1000 µl,
equipped with aerosol filter
Samples
Sample capacity
Up to 16 preparations in parallel, incl. handling of individual
samples
Processing of different source materials possible with identical
pipetting protocols
Sample preparation
Insertion of the pipette tips, samples and elution tubes into the
sample tray
Insertion of pre-filled reagent strips/plates into the sample tray
Sample volume
Up to 10 ml or up to 180 mg