BALLOON INFLATION AND DEFLATION:
1.
 To inflate the balloon, squeeze the trigger allowing the plunger to return 
  to the resting position (0 ATM/BAR). Release grip on the trigger, locking 
  the plunger into position.
2. To increase pressure, rotate the handle clockwise until the desired 
  inflation pressure is reached. The locking mechanism maintains the 
 pressure.
CAUTION:  Do not exceed 12 ATM.
NOTE: Loss of pressure may indicate a leak in the system.
3. To deflate the balloon (rotate the handle counter-clockwise to relieve 
  pressures above 6 ATM) squeeze the trigger and pull the handle back to 
  generate the desired negative pressure. Release the grip on the trigger 
  to lock the plunger in the negative pressure position.

STOrAgE: Store in a cool, dry place.

WArrANTY

The manufacturer warrants that reasonable care has been used in the design and manufacture 
of this device.  This warranty is exclusive and manufacturer makes no other representations or 
warranties of any kind to customers, its end users, or to any third parties with respect to the device 
and hereby expressly disclaims any and all other warranties, express or implied, statutory or 
otherwise, including, but not limited to, infringement and the implied warranties of merchantability 
and fitness for a particular purpose, even if manufacturer is aware of such purpose.  Handling and 
storage of this device, as well as other factors relating to the patient, diagnosis, treatment, implant 
procedures, and other matters beyond the control of the manufacturer, directly affect the device 
and the results obtained from its use.  The manufacturer’s obligation under this warranty is limited 
to the replacement of the device.  Under no circumstances shall manufacturer be liable to customer 
or any other person or entity for any punitive, special, incidental or consequential damages directly 
or indirectly arising from the use of this device.  The manufacturer neither assumes, nor authorizes 
any other person to assume for it, any other or additional liability or responsibility in connection 
with this device.  This warranty shall not apply, and manufacturer assumes no liability with respect 
to, devices that have been (i) modified, changed, altered, misused, mishandled, repaired, reused, 
reprocessed, refurbished or resterilized; (ii) subjected to improper maintenance, testing or storage, 
accident, tampering, or inadequate protection against shock, vibration, excessively high or low 
temperatures, overpressure, or physical, environmental or electrical stress; (iii) been used outside 
the approved “Indications for Use” as cleared by the relevant competent authority, used contrary 
to the use outlined in the device specifications, or in an application or environment for which such 
device was not designed or contemplated; or (iv) distributed or used contrary to applicable federal, 
state, local or regulatory standards.  

rX only: CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the 
order of a physician.

 
Single Use

Caution: Consult accompanying document

Sterile if package is unopened or undamaged
Non-pyrogenic

 

Manufacturer: 

 

                                   www.merit.com

 

Merit Medical Systems, Inc.  South Jordan, Utah  84095  U.S.A. 

 

1-801-253-1600  

 

U.S.A. Customer Service 1-800-356-3748

 

Authorized Representative:

 

Merit Medical Ireland Ltd, Parkmore Business Park West, 

 

Galway, Ireland

401997002/C    ID 031513

60ml Inflation Device

I N S T r U C T I O N S   F O r   U S E

DESCrIPTION:
The BIG60

®

 Inflation Device is a 60ml disposable inflation device capable of producing 

a maximum pressure of 12 ATM/BAR, fitted with a threaded plunger assembly with 
a lock/release bar, a flexible high pressure tube, and a three-way medium pressure 
stopcock.  The BIG60 Inflation Device is designed to generate and monitor inflation 
pressures over a range of -.68 to 12ATM/BAR (-10 to 176 psi) with an accuracy of 3% 
ATM/BAR.

INTENDED USE:
This inflation device is used to inflate and deflate an angioplasty balloon or other 
interventional device, and to measure the pressure within the balloon.

PrECAUTIONS:
•  Inspect the BIG60

 

 Inflation Device and packaging for damage prior 

  to use. Do not use product if opened or damaged. Confirm the product is 
  consistent with the package label. Contact Customer Service to report 
  and replace damaged product.
•  For single patient use only. Do not reuse, reprocess, or resterilize. Reuse, 
  reprocessing, or resterilization may compromise the structural integrity 
  of the device and/or lead to device failure which, in turn, may  result in 
  patient injury, illness, or death. Reuse, reprocessing, or resterilization 
  may also create a risk of contamination of the product and/or cause 
  patient infection or cross-infection, including, but not limited to, 
  the transmission of infectious disease(s) from one patient to another. 
  Contamination of the product may lead to injury, illness, or death of 
  the patient.
•  Graduations on device barrel are for reference only and are not intended 
  for exact measurement.

INSTrUCTIONS FOr USE:
1.
 Prior to use, free the plunger tip by twisting the device plunger/handle    
360 degrees clockwise.
2. To prep device, simply aspirate up to 60ml of contrast solution into the    
syringe by pulling back on the device handle.
CAUTION:  Inspect the device tubing and stopcock (if used) to ensure there is no 
air in the system.
3. Remove any excess air by orienting the device upwards, squeezing the    
trigger located on the device handle and pushing the handle forward.  

 

Push forward until the device plunger tip is oriented with the black  

 

arrow printed on the device barrel (approximately 35ml).

ATTACHINg THE INFLATION DEVICE TO THE BALLOON:
NOTE:
 Refer to the manufacturer’s directions accompanying the balloon dilation 
catheter, or other interventional device, for specific information on use, maximum 
inflation pressure, precautions, and warnings for that device.
1. Prepare and test the balloon catheter according to the catheter  

 

  manufacturer’s instructions for use.
2. Create a fluid-fluid connection between the balloon and the inflation 
  device extension tube and connect the luer connectors securely.
3. Squeeze the trigger and pull back on the handle to apply vacuum to the 
 balloon.