background image

1. Keep body face clean, prevent liquid material from getting into slider
to destroy electronics.
2. Face should be cleaned with a clean, dry, lint-free cloth. Lubricate
body with a few drops of clock oil. Acetone & alcohol must not be used.
3. Keep battery compartment clean and free of corrosion.

PROBLEM: 

SOLUTION:

 

Every second 5 digits 

Battery voltage is under 1.45V. Remove 

jump simultaneously 

cover and replace battery.

 

Display cannot count 

Faulty circuit. Remove battery, after 30 
sec. put battery back into compartment.

 

Display shows 000.00 

Function buttons and sliders signal end 

or IN 00.000   

may have shorted out. Remove the 
cover, put the button springs in place

 

with rubber covers. Contact must be 
unobstructed.

 

Function buttons not active  

Springs or rubber covers may be out of 
shape due to excessive pressing. See   
solution above for tips.

 

Reading error for full 

Sensor may have dirt or deposits in it 

length is 

 

0.1mm 

Remove 

cover and slider assembly.

 

Blow off sensor face with clean

 

pressurized air (

5kg/cm squared),

 

clean with a dry, lint-free cloth.

 

No display on LCD screen 

1.) Poor battery contact. Check battery

 

compartment, clean if necessary.

 

2.) Battery voltage is under 1.3V.

 

Replace with correct battery.

 

 

Digital Caliper 

E T 3 5 6 0

O W N E R ' S   M A N U A L

CARE & MAINTENANCE

 

TROUBLESHOOTING

 

In accordance with the 

    European Directive 

ROHS 2002/95/CE 

D4 06 07 60503 001

FOR YOUR SAFETY, and
CARE OF THIS TOOL

Please read these instructions 
carefully and retain them for future use. 

03.04.0487ECC