Users’ Manual 

 

EAN-26CCJ Inclinometer calibration check jig  

  

  

 

Page

 | 

1

 

1  INTRODUCTION 

The calibration check jig is used to ensure that the original factory calibration of the inclinometer probe is 
unchanged. As a good practice, the calibration of the inclinometer probe must be checked periodically with 
this jig, preferably once every 1~3 months depending on the usage of probe/criticality of site or in case of 
uncertainty.  

For checking, inclinometer probe is inserted in the wall mounted calibration check jig and readings for 0

°

±

 14

°

 and 

±

 28

o

 positions are noted down. The initial readings are taken as the reference readings for future 

reference. Later on when the probe’s calibration is to be checked, the readings are again taken at 0

°

, +14

°

+ 28

°

, - 14

°

, -28

°

 and compared with the initial reference reading. If any drift is detected in the probe 

readings at a later date, it indicates that the probe needs calibration. 