Diagnostic variables mnemonics
Table 8-2:
AbS
Absolute conductivity (µS/cm or mS/cm)
0 A ir
Sensor zero in air
CELL ConSt
Sensor cell constant (used in C mode)
tSLOPE
Temperature slope in %/° C
CAL F
Calibration factor
SoFt
Software version
HArd
Hardware version
FAuLtS
Show fault messages
nonE
No fault messages in memory
8.2.2
Fault conditions
Three classes of error conditions/problems are detected and differentiated between by the
diagnostic program. System disabling problems are faults caused by failures in the loop or
significant variations in the process. System non-disabling problems are warnings and deal
with input or A to D conversion settings. The third class of detection problems are error
messages and occur when the calibration limits are exceeded.
Disabling faults
The following things happen when a disabling fault occurs.
1.
Both FAULT and HOLD annunciation fields become active (
).
2.
The process variable flashes at the rate of 1 second on and 1 second off.
3.
The appropriate fault message alternates with the normal Temperature/Current
output display (see
).
4.
The output current loop is forced to run the non-zero fault value last entered or held
at last value if fault value = 0 if the transmitter is not in the TEST, HOLD, or Multidrop
operational modes.
5.
A 0-1 mA output signal is available for external use when system disability
conditions are active. These conditions drive this output to 1 mA. Please contact
factory for specific applciation information.
Diagnostics and troubleshooting
82
Rosemount 5081
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...