Message
Explanation
See section
bAd rtd
Bad temperature reading
bAd rtd, TEMP HI, TEMP LO, and
rtd OPEn
TEMP HI
Temperature reading exceeds 150 °C (302 °F)
bAd rtd, TEMP HI, TEMP LO, and
rtd OPEn
TEMP LO
Temperature reading is less than -15 °C (5 °F)
bAd rtd, TEMP HI, TEMP LO, and
rtd OPEn
rtd OPEn
RTD or thermistor is open
bAd rtd, TEMP HI, TEMP LO, and
rtd OPEn
SenSE OPEn
Sense line is not connected
PH in
Raw millivolt reading from pH sensor is too
large
SLOPE HI
pH sensor slope exceeds 62 mV/pH
SLOPE LO
pH sensor slope is less than 40 mV/pH
-0- OFFSEt
Zero offset during standardization exceeds
programmed limit
GLASSFAIL
Measured glass impedance is less than pro-
grammed limit
FACt FAIL
Unit has not been factory calibrated
CPU FAIL
Internal CPU tests have failed
9.4.2.13
ROM FAIL
Internal memory has failed
9.4.2.13
AdC
Analog to digital conversion failed
bAd Gnd
Bad ground
In too biG
mV signal from pH sensor is too large
RitE Err
CPU PCB jumper (JP-1) has been removed
bAd SEnSor
Bad sensor means that the sensor current is a large negative number.
bAd SEnSor may appear for a while when the sensor is first placed in service. Observe the
sensor current (go to SEnSor Cur under the Diagnostic menu). If the sensor current is
moving in the positive direction, there is probably nothing wrong, and the error message
should soon disappear.
Procedure
1.
Verify that wiring is correct. Pay particular attention to the anode and cathode
connections.
2.
Verify that the transmitter is configured for the correct measurement.
Configuring the measurement sets (among other things) the polarizing voltage.
Applying the wrong polarizing voltage to the sensor can cause a negative current.
3.
Replace the sensor membrane and electrolyte solution and clean the cathode if
necessary.
Diagnostics and troubleshooting
Instruction Manual
95
Содержание Rosemount5081
Страница 1: ...Instruction Manual LIQ MAN 5081 Rev A September 2017 Rosemount 5081 Explosion Proof Transmitter ...
Страница 4: ......
Страница 10: ...Contents vi Rosemount 5081 ...
Страница 12: ...Startup procedure 2 Rosemount 5081 ...
Страница 23: ...Mounting 5081 on a pipe Figure 3 2 Installation Instruction Manual 13 ...
Страница 24: ...Installation 14 Rosemount 5081 ...
Страница 28: ...Wiring 18 Rosemount 5081 ...
Страница 50: ...Programming basics 40 Rosemount 5081 ...
Страница 130: ...Diagnostics and troubleshooting 120 Rosemount 5081 ...
Страница 136: ...Digital communications 126 Rosemount 5081 ...
Страница 138: ...NOTICE Consult the factory for additional information regarding service or repair Return of material 128 Rosemount 5081 ...
Страница 140: ...Engineering drawings 130 Rosemount 5081 ...
Страница 141: ...Engineering drawings Instruction Manual 131 ...
Страница 142: ...Engineering drawings 132 Rosemount 5081 ...
Страница 143: ...Engineering drawings Instruction Manual 133 ...
Страница 144: ...Engineering drawings 134 Rosemount 5081 ...
Страница 145: ...Engineering drawings Instruction Manual 135 ...
Страница 146: ...Engineering drawings 136 Rosemount 5081 ...
Страница 151: ...EU Declarations of Conformity 138 Rosemount 5081 ...