7.6.3.1 Locating the Sensor
Fig. 7-12: Location of the EO
2
Sensor Unit
Basically X-STREAM analyzers provide dif-
ferent variations of internal designs:
• In instruments with internal heated box
covering the physical components, elec-
trochemical sensors are installed outside
this box.
Instruments without internal thermostatic
control have the sensor installed onto
the basic mounting plate
(see left side of
figure 7-12)
.
•
XEGP may also have the electrochemical
oxygen sensor installed at the rear panel
(see right side of figure 7-12).
If your analyzer features eO2 sensor at rear
panel, continue with page 7-57.
7.6 Replacing Worn Out Sensors
1 eO2 sensor Unit
2 Cover for rear panel installation
1
1 (shown behind transparently
visualized cover)
2
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......