7.4.6 Verifying a Calibration
7.4.6 Verifying a Calibration
Starting from the MEASUREMENT SCREEN
press
down
to open the MAIN MENU and
enter
CONTROL
.
Enter APPLY GAS
For instruments
without
internal and/or ex-
ternal valves simply apply either span or zero
calibration gas to the sample gas inlet. If the
calibration still is proper, the reading on the
MEASUREMENT SCREEN should show the
related value.
For instruments
with
internal and/or external
valves follow the procedure below:
Multi-channel unit:
Select the component to be verified in
SELECT COMPONENT.
Changing the "
Applied gas
" parameter opens
the related valve.
Available options:
SpanGas-1 ... -4, ZeroGas, SampleGas,
Blowback, All closed.
"
Flow"
shows the current gas flow, while "
Con-
centration"
should show the expected value,
if the calibration is valid and correct.
Zero calibration..
Span calibration..
Advanced calibration..
Apply gas..
Lock menus!
Acknowledgements
Pump 1:
Off
Pump 2:
Off
Ranges..
Ch1
Apply gas
Applied gas:
Span gas
Flow
1.00 l/min
Concentration
25.000 ppm
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......