6.2.3 Setup Menu
6.2.3.6 Setup Alarms
Setup..
Alarms..
Alarms
Alarms Output Failsafe:
Yes
Concentration..
Conc Average..
Temperature..
Pressure..
Flow..
Alarms 2
Calculator A..
Calculator B..
Calculator C..
Calculator D..
This menu and its submenus enable to con
-
figure a couple of alarm conditions.
In case an alarm goes off, a status is set and
the related pictogram shows in the display.
Also digital outputs may be configured to be
activated (
Configure concentration alarms (
6-81 )
Configure alarms for concentration averages
(
Configure temperature alarms (
Configure pressure alarms (
Configure flow alarms (
X-STREAM XE analyzers can be upgraded
with optional software packages, to provide
four calculators (A ... D).
Their results can be monitored to activate
alarms (
If the system is setup accordingly, the access
code for level 3 must be entered to gain ac-
cess to this menu.
Select if the alarms outputs are failsafe (
Yes
)
or not (
No
): Failsafe means, relay output coils
are powered during normal operation.
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......