Signal:
Comp1
OutRange:
0-20 mA
Low Scale:
0.00
Max Scale:
100.00
AutoScale:
Yes
FailMode:
Live
0/4 mA:
0.00
20 mA:
100.00
Hold:
No
6.2.3.4.1.1
Setup Analog Outputn
6.2.3 Setup Menu
"
Signal
” specifies the value associated with
the selected output. The following options
(partly dependent on the number of measu
-
ring channels and sensors installed) are
available:
Signal
*)
Description
None
The analog output is deactivated
0 mA
A 0 or 4 mA signal is generated, e.g. to check the signal processing in an
external system. Whether a 0 or 4 mA signal is generated, is set by the "
Out
range
” line (
next page).
20 mA
A 20 mA signal is generated, e.g. to check the signal processing in an exter-
nal system.
Comp1 ... 5
Gas concentration
Temp1 ... 5
Temperature
Press1 ... 5
Pressure
Flow1 ... 5
Flow
Calc A ... D
Result of calculator
RawVal1 ... 5
Raw value
RangeID1 ... 5 ID of selected range
*) Numbers 1 to 5 refer to components [chan
-
nels] 1 to 5: In case of secondary measure-
ments, this means, the selected value is that
of the sensor assigned to the given compo-
nent (Press2 is the pressure value of the
sensor assigned to component 2).
In contrast, capital letters A to D imply that
these calculator results are component
[channel] independent (Calc C is the result
of calculator C).
Setup..
In-/Outputs..
Analog outputs..
Output
1...5
..
Tab. 6-1: Analog Output Signals
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......