Emerson Process Management GmbH & Co. OHG
4-22
X-STREAM XE
Instruction Manual
HASXEE-IM-HS
10/2012
4.6.2 Installation - X-STREAM XEXF Field Housings
Power and signal cables are connected using
internal screw-type terminals. This requires
opening the unit by releasing the fasteners
on the housing.
Gas connectors are accessible from the un
-
derside of the instrument.
The number and configuration of the gas
inlets and outlets depends on the analytical
application, and is noted on a sticker on the
underside of the instrument next to the con-
nectors.
To simplify installation, we recommend label
-
ling the gas lines in accordance with these
markings. This avoids confusion should the
analyzer need to be disconnected for main
-
tenance.
Fig. 4-17: X-STREAM XEXF Field Housings - Terminals, Cable Glands and Gas Fittings
5 Glands for signal cables
6 Gas inlets and outlets
7 Plugs for openings to connect housings
8 Ethernet connectors
1 Terminals for signal cables
2 Mains filter
3 Power connections with integrated fuses
4 Gland for power cable
6
5
7
Gas inlets
Gas outlets
1
8
4
2
3
Note on XDF!
In case of the dual compartment version XDF,
the electrical connections are established in the
upper compartment, and the gas connections to
fittings at the lower compartment.
Besides this, the design and layout of terminals
and fittings are the same as with the single com
-
partment version XEF.
GASKETS AT LOW
TEMPERATURES
Consider that enclosure gaskets
may be frozen if the instrument is
installed outdoors. Carefully open
the enclosure at temperatures
below -10 °C / 14 °F to not damage
the gaskets.
Damaged gaskets void the ingress
protection, possibly causing pro-
perty damage, personal injury or
death.
Содержание Rosemount X-Stream Series
Страница 370: ...Emerson Process Management GmbH Co OHG 7 82 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 414: ...Emerson Process Management GmbH Co OHG 8 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 486: ...Emerson Process Management GmbH Co OHG A 4 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 488: ...Emerson Process Management GmbH Co OHG A 6 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 490: ...Emerson Process Management GmbH Co OHG A 8 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 A 1 Modbus Specification ...
Страница 526: ...Emerson Process Management GmbH Co OHG A 44 X STREAM XE Instruction Manual HASXEE IM HS 10 2012 ...
Страница 535: ......