background image

6

Troubleshooting

This section provides information for basic troubleshooting. See 

Wireless

Pressure Gauge Reference Manual

 for advanced troubleshooting.

Device status

The flashing LED indicates device status using the colors descibed.

Table 6-1: Status Descriptions

LED color

Device status

 

Green

Functioning properly

 

Amber Battery is low, battery replacement recommended

 

Red

Battery replacement required OR Device is
malfunctioning

 

No
color

No power, verify ON/OFF switch is in “on” position

6.1

Pressure measurement

If the mounting effects have not been eliminated after completing 

Eliminate

mounting effects

, perform this alternative procedure for verifying the

pressure value.

Procedure

1. From the 

HOME

 screen, eter the HART

®

 Fast Key sequence.

Device dashboard Fast
Keys

2, 2, 1, 1, 1

2. Follow the commands to perform the procedure.

6.2

Wireless connectivity

If the device has not joined to the network after power up, verify the
following:

Active Advertising has been enabled on the Smart Wireless Gateway

Quick Start Guide

Feburary 2021

14

Emerson.com/Rosemount

Содержание Rosemount Wireless Pressure Gauge

Страница 1: ...Quick Start Guide 00825 0100 4045 Rev CA Feburary 2021 Rosemount Wireless Pressure Gauge with Wireless HART Protocol ...

Страница 2: ...nvironment must be in accordance with appropriate local national and international standards codes and practices Ensure device is installed in accordance with intrinsically safe or non incendive field practices Electrical shock could cause death or serious injury Care must be taken during transportation of device to prevent electrostatic charge build up Device must be installed to ensure a minimum...

Страница 3: ...ice check 8 Optional Normal range indication option 9 Installation procedure 10 Troubleshooting 14 Product Certifications 16 Declaration of conformity 21 China RoHS 24 Feburary 2021 Quick Start Guide Quick Start Guide 3 ...

Страница 4: ...Quick Start Guide Feburary 2021 4 Emerson com Rosemount ...

Страница 5: ...i seize paste or PTFE tape for NPT threaded connection Standard tools e g screwdriver wrench pliers AMS AMS Wireless Configurator version 12 0 or later or Field Communicator Feburary 2021 Quick Start Guide Quick Start Guide 5 ...

Страница 6: ...k Start Guide The following options are also available and will ship with the Rosemount Wireless Pressure Gauge if ordered Rosemount 306 Integral Manifold Model code S5 B4 Bracket Model code B4 Quick Start Guide Feburary 2021 6 Emerson com Rosemount ...

Страница 7: ...erification Cert QP Calibration Cert and Tamper Evident Seal Q8 Material Traceability cert per EN 102043 1 Q15 Cert of Compliance to NACE MR0175 ISO 15156 for wetted materials Q25 Cert of Compliance to NACE MR00103 for wetted materials Normal Range Indication Model code LK Feburary 2021 Quick Start Guide Quick Start Guide 7 ...

Страница 8: ...ed to be installation ready To check device battery prior to installation perform the following Procedure 1 Perform Turn on device 2 Slide the ON OFF switch to the OFF position until ready for use Quick Start Guide Feburary 2021 8 Emerson com Rosemount ...

Страница 9: ...nics assembly is connected to the needle 5 Remove any debris on the dial so it does not become trapped under the sticker 6 Peel back the white paper backing of the sticker 7 Slowly lower the sticker onto the surface of the dial in the desired location and rub it in place firmly Repeat steps 6 and 7 until desired indication locations are set Stickers should be applied in an environment where the am...

Страница 10: ...port atmospheric reference on the process pressure gauge is located in the neck of the device behind the housing The vent path is between the housing and sensor CAUTION Keep the vent path free of any obstruction including but not limited to paint dust and lubrication by mounting the device so the process can drain away Quick Start Guide Feburary 2021 10 Emerson com Rosemount ...

Страница 11: ...king properly Procedure 1 Twist the cover counterclockwise to remove it 2 Slide the OFF ON switch to the ON position to initiate the power sequence Note During the power sequence the dial tests full range of motion and LED flashes amber 3 Once the power sequence ends verify the LED flashes green Feburary 2021 Quick Start Guide Quick Start Guide 11 ...

Страница 12: ...the HART symbol 5 5 AMS Wireless Configurator Procedure 1 Start AMS Wireless Configurator 2 From the View menu select Device Connection View 3 Double click the device under the HART modem 5 6 Eliminate mounting effects Devices are factory calibrated Once installed it is recommended to perform this step to eliminate potential error caused by mounting position or static pressure Instructions for usi...

Страница 13: ...operly toggling off and on reduces battery life Note If Network ID and Join Key is specified at order entry then the device will automatically search and connect to the wireless network when powered on 5 8 Join device to network Procedure 1 Obtain Network ID and Join Key for the wireless network available in wireless gateway 2 From the HOME screen enter the HART Fast Key sequence Device dashboard ...

Страница 14: ...ning No color No power verify ON OFF switch is in on position 6 1 Pressure measurement If the mounting effects have not been eliminated after completing Eliminate mounting effects perform this alternative procedure for verifying the pressure value Procedure 1 From the HOME screen eter the HART Fast Key sequence Device dashboard Fast Keys 2 2 1 1 1 2 Follow the commands to perform the procedure 6 2...

Страница 15: ...ice match the Network ID and Join Key of the Gateway The Network ID and Join Key may be obtained from the Smart Wireless Gateway on the Setup Network Settings page on the web interface Feburary 2021 Quick Start Guide Quick Start Guide 15 ...

Страница 16: ...he following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device Changes or modification to the equipment not expressly approved by Rosemount Inc could void the user s authority to operate the equipment Cet appareil est conforme à la Partie 15 de la réglementation FCC Son fonc...

Страница 17: ...2011 FM 3610 2010 UL Standard 50 Eleventh Edition UL 61010 1 3rd Edition ANSI ISA 60079 0 12 00 01 2013 ANSI ISA 60079 11 12 02 01 2013 ANSI IEC 60529 2004 Markings IS CL I DIV 1 GP A B C D T4 Class 1 Zone 0 AEx ia IIC T4 Ga T4 40 C Ta 70 C when installed per Rosemount drawing 00G45 1020 Type 4X IP66 67 Special Conditions for Safe Use X 1 Do not replace battery when explosive atmosphere is present...

Страница 18: ...er ou nettoyer avec des produits solvants ou un chiffon sec 4 Substitution of components may impair intrinsic safety La substitution de composants peut compromettre la sécurité intrinsèque 7 3 Europe I1 ATEX Intrinsic Safety Certificate Baseefa16ATEX0005X Standards EN 60079 0 2012 A11 2013 EN 60079 11 2012 Markings II 1 G Ex ia IIC T4 Ga T4 40 C Ta 70 C IP66 67 Special Conditions for Safe Use X 1 ...

Страница 19: ...grated into a system where the process connection is not grounded 3 Do not change the battery when an explosive atmosphere is present 4 Only replace battery with Rosemount Part No 00G45 9000 0001 7 5 Brazil I2 INMETRO Intrinsic Safety Certificate UL BR 16 0826X Standards ABNT NBR IEC 60079 0 2008 Errata 1 2011 ABNT NBR IEC 60079 11 2009 Markings Ex ia IIC T4 Ga T4 40 C Ta 70 C Special Condition fo...

Страница 20: ...x ia IIC T4 Ga X T4 40 C Ta 70 C IP66 67 Special Condition for Safe Use X See certificate for special condition 7 8 Korea IF KTL Intrinsic Safety Certificate 16 KA4BO 0540X Markings Ex ia IIC T4 Ga T4 40 C Ta 70 C Special Condition for Safe Use X See certificate for special condition Quick Start Guide Feburary 2021 20 Emerson com Rosemount ...

Страница 21: ...8 Declaration of conformity Feburary 2021 Quick Start Guide Quick Start Guide 21 ...

Страница 22: ...Quick Start Guide Feburary 2021 22 Emerson com Rosemount ...

Страница 23: ...Feburary 2021 Quick Start Guide Quick Start Guide 23 ...

Страница 24: ...9 China RoHS Quick Start Guide Feburary 2021 24 Emerson com Rosemount ...

Страница 25: ...Feburary 2021 Quick Start Guide Quick Start Guide 25 ...

Страница 26: ...Quick Start Guide Feburary 2021 26 Emerson com Rosemount ...

Страница 27: ...Feburary 2021 Quick Start Guide Quick Start Guide 27 ...

Страница 28: ...erson com 2021 Emerson All rights reserved Emerson Terms and Conditions of Sale are available upon request The Emerson logo is a trademark and service mark of Emerson Electric Co Rosemount is a mark of one of the Emerson family of companies All other marks are the property of their respective owners ...

Отзывы: