Quick Start Guide
00825-0100-4145, Rev BA
February 2018
Rosemount
™
Smart Pressure Gauge
Страница 1: ...Quick Start Guide 00825 0100 4145 Rev BA February 2018 Rosemount Smart Pressure Gauge ...
Страница 2: ...y other local requirements Consult current regulations and requirements before shipping Explosions could result in death or serious injury Installation of device in an explosive environment must be in accordance with appropriate local national and international standards codes and practices Ensure device is installed in accordance with intrinsically safe or non incendive field practices Electrical...
Страница 3: ...el code S5 B4 bracket model code B4 Rosemount 1199 Seal model code S1 Product Certification model codes below Q4 Calibration Cert QG Calibration Cert and GOST Verification Cert QP Calibration Cert and Tamper Evident Seal Q8 Material Traceability cert per EN 102043 1 Q15 Cert of Compliance to NACE MR0175 ISO 15156 for wetted materials Q25 Cert of Compliance to NACE MR00103 for wetted materials Norm...
Страница 4: ...tickers to desired size 2 Remove housing cover 3 Slide ON OFF switch to the OFF position and wait for LED to stop flashing 4 Gently move the needle in the clockwise direction until it is pointing at the Red X Note Use caution as the electronics assembly is connected to the needle 5 Remove any debris on the dial so it does not become trapped under the sticker 6 Peel back the white paper backing of ...
Страница 5: ...pressure port atmospheric reference is located in the neck of the device behind the housing The vent path is between the housing and sensor see Figure 1 Figure 1 Low Side Pressure Port A Low side pressure port atmospheric reference Keep the vent path free of any obstruction including but not limited to paint dust and lubrication by mounting the device so the process can drain away OR A ...
Страница 6: ... flashes amber 3 Once the power sequence ends verify the LED flashes green Note The LED may display several colors see Figure 1 in Troubleshooting on page 7 for device statuses Step 4 Connect to device Field Communicator 1 Turn on the Field Communicator 2 From the Main menu select the HART symbol AMS Configurator 1 Start AMS Configurator 2 From the View menu select Device Connection View 3 Double ...
Страница 7: ...shooting This section provides information for basic troubleshooting See the Reference Manual for advanced troubleshooting Device status The flashing LED indicates device status using the colors descibed in Table 1 Table 1 Status Descriptions Pressure measurement If the mounting effects have not been eliminated after completing Step 5 perform this alternative procedure for verifying the pressure v...
Страница 8: ...assification gas and temperature class This information is clearly defined in the respective codes 5 4 USA I5 U S A Intrinsically Safe IS Certificate CSA 70047656 Standards FM 3600 2011 FM 3610 2010 UL Standard 50 Eleventh Edition UL 61010 1 3rd Edition ANSI ISA 60079 0 12 00 01 2013 ANSI ISA 60079 11 12 02 01 2013 ANSI IEC 60529 2004 Markings IS CL I DIV 1 GP A B C D T4 Class 1 Zone 0 AEx ia IIC ...
Страница 9: ...79 11 2012 Markings II 1 G Ex ia IIC T4 Ga T4 40 C Ta 70 C IP66 67 Special Conditions for Safe Use X 1 The plastic enclosure may constitute a potential electrostatic ignition risk and must not be rubbed or cleaned with a dry cloth 2 The measured capacitance between the equipment enclosure and metallic inline sensor module is 4 7pF This must be considered only when the SPG is integrated into a syst...
Страница 10: ...nc 8200 Market Boulevard Chanhassen MN 55317 9685 USA to which this declaration relates is in conformity with the provisions of the European Community Directives including the latest amendments as shown in the attached schedule Assumption of conformity is based on the application of the harmonized standards and when applicable or required a European Community notified body certification as shown i...
Страница 11: ... WPG Wireless Pressure Gauge only Harmonized Standards EN 300 328 V2 1 1 EN 301 489 1 V2 2 0 EN 301 489 17 V3 2 0 EN 61010 1 2010 EN 62479 2010 ATEX Directive 2014 34 EU Models WPG SPG Baseefa16ATEX0005X Intrinsic Safety Certificate Equipment Group II Category 1 G Ex ia IIC T4 Ga T4 7D Harmonized Standards EN 60079 0 2012 A11 2013 EN 60079 11 2012 Quick Start Guide 11 February 2018 ...
Страница 12: ... of 3 ATEX Notified Bodies Baseefa Notified Body Number 1180 Rockhead Business Park Staden Lane Buxton Derbyshire SK17 9RZ United Kingdom ATEX Notified Body for Quality Assurance Baseefa Notified Body Number 1180 Rockhead Business Park Staden Lane Buxton Derbyshire SK17 9RZ United Kingdom ...
Страница 13: ...O O ༣փ㓴Ԧ Housing Assembly O O O O O O Րᝏಘ㓴Ԧ Sensor Assembly X O O O O O 㓴Ԧ Battery Assembly X O O O O O ᵜ㺘Ṭ ᦞ SJ T11364Ⲵ㿴ᇊ㘼ࡦ This table is proposed in accordance with the provision of SJ T11364 O Ѫ䈕䜘ԦⲴᡰᴹ 䍘ᶀᯉѝ䈕ᴹᇣ 䍘Ⲵਜ਼䟿 վҾGB T 26572ᡰ㿴ᇊⲴ䲀䟿㾱 O Indicate that said hazardous substance in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Ѫ 䈕䜘Ԧᡰ Ⲵᡰᴹ 䍘ᶀᯉ䟼ˈ㠣ቁᴹа 䍘ᶀᯉѝ...
Страница 14: ...t Emerson Terms and Conditions of Sale are available upon request The Emerson logo is a trademark and service mark of Emerson Electric Co Rosemount is a mark of one of the Emerson family of companies All other marks are the property of their respective owners 2018 Emerson All rights reserved Europe Regional Office Emerson Automation Solutions Neuhofstrasse 19a P O Box 1046 CH 6340 Baar Switzerland...