Section 2:
Installation
April 2017
Instruction Manual
IM-106-880C, OI
30
Pneumatic
Installation
2.5
Pneumatic Installation
Pneumatic system connections depend on whether reference air set, calibration solenoids, and/
or blowback equipment options are equipped on your transmitter. Refer to the following
paragraphs and select the option that applies to your transmitter configuration.
2.5.1
Reference Air Set Option (only)
When no options or only the reference air set option is equipped, use the following procedure to
install the pneumatic system components.
1.
Refer to
. Connect the reference air set (regulator/filter and pressure gage) to the instrument air
inlet on the electronics housing and to the inlet side of the dilution air flow meter.
2.
Connect the dilution air flow meter output to the dilution air inlet fitting on the sensor housing.
3.
Install an air line between the instrument air outlet fitting on the electronics housing and the tee fitting on
the sensor housing.
Do not use 100% nitrogen as an O
2
low gas. It is suggested that O
2
low gas be between
0.4% and 2.0% O
2
. Do not use gases with hydrocarbon concentrations of more than 40
parts per million. Failure to use proper gases will result in erroneous readings.
4.
One CO gas and two O
2
gases are used to calibrate the Hazardous Area OCX 8800:
•
CO - 1000 ppm or up to 4%, Balance air
•
O
2
low gas - 0.4% , Balance N
2
•
O
2
high gas - 8%, Balance N
2
5.
Connect the output of the test gas sources to the inlet port of the CAL GAS flow meter. Install an air line
between the flow meter outlet port and the CAL GAS inlet fitting on the sensor housing.
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......