Instruction Manual
IM-106-880C, OI
Repair Sensor
Housing
April 2017
Repair Sensor
Housing
18
9
5.
Connect the heater cable leads to the heater power connector (J3).
6.
Connect the line (L1 wire) to the L1 terminal, and the neutral (N wire) to the N terminal
on the AC power input terminal block,
7.
Connect the ground lead to the ground stud. Secure the connection with two nuts.
Attach a separate ground lead (G wire) from the ground stud to the G terminal on the
power input terminal block.
8.
If used, connect external relay leads to the alarm output relay terminal.
9.
Connect the test gas and instrument air lines to the remote electronics housing.
Connect the calibration gas line and instrument air line to the remote electronics
housing.
10.
Refer to
and
. Make sure all test gas lines and electrical
connections are complete.
11.
Install the cover on the electronics housing.
12.
Turn on the test gasses at the cylinders and open the instrument air valve.
13.
Restore power to the system.
9.4
Repair Sensor Housing
Use the following procedures to remove damaged components from the Hazardous Area OCX
8800 sensor housing and to install new replacement parts. Disassemble the unit only as needed
to replace damaged components. Use the assembly procedures that apply to install
replacement parts and reassemble the unit.
9.4.1
Sensor Housing Disassembly
Remove Cover and Terminals Insulator
1.
Loosen screw (1,
) and slide locking clip (2) away from cover. Retighten screw
(1).
2.
With two hands or strap wrench, turn cover (3) counterclockwise to loosen. Unthread
and remove cover.
3.
Inspect cover o-ring (4) for wear or damage. Replace cover o-ring if damaged.
4.
Unsnap terminal marking plates (5) and remove terminal insulator (6).
Содержание Rosemount OCX 8800
Страница 2: ......
Страница 41: ...Instruction Manual IM 106 880C OI Section 2 Installation April 2017 Section contents 17 ...
Страница 192: ...Section 7 Foundation Fieldbus April 2017 Instruction Manual IM 106 880C OI 16 6 About AMS and DeltaV Software ...
Страница 205: ...Alarm Relay Events April 2017 Instruction Manual IM 106 880C OI 17 8 Alarm Relay Events ...
Страница 251: ...Replace Tube Fittings April 2017 Instruction Manual IM 106 880C OI 22 4 Install Tube Fittings ...
Страница 259: ...Section 10 Replacement Parts April 2017 Instruction Manual IM 106 880C OI 23 2 O2 Cell and Heater Strut Assembly ...
Страница 261: ...Instruction Manual IM 106 880C OI Safety Data April 2017 Appendix contents 233 ...
Страница 263: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 235 ...
Страница 267: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 239 ...
Страница 271: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 243 ...
Страница 273: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 245 ...
Страница 275: ...Instruction Manual Safety Data IM 106 880C OI April 2017 IMPORTANT Safety Instructions 247 ...
Страница 277: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 249 ...
Страница 279: ...Instruction Manual Safety Data IM 106 880C OI April 2017 Safety Instructions 251 ...
Страница 298: ...SPA with HART Alarm April 2017 Instruction Manual IM 106 880C OI 27 2 Configuration Calibratio n ...
Страница 300: ...Instruction Manual IM 106 880C OI Index April 2017 Manual index 27 4 ...
Страница 304: ...Index April 2017 Instruction Manual IM 106 880C OI 27 8 Manual index ...
Страница 305: ......